2.7L Extractor System
Lab Scale Supercritical CO2
Dimensions
- Width: 5.9 feet (1.8 meters)
- Height: 6 feet (1.8 meters)
- Depth: 4.6 feet (1.4 meters)
- Weight: 1543 lbs (700kg)
- Recommended Room Size: 9.8 x 9.8 feet (3 x 3 meters) with a ceiling height of 8.5 feet (2.6 meters)
Technical Specifications
- Voltage: 460 AC-V or 400 AC-V
- Frequency: 60 Hz or 50 Hz
- Installed Power: 4.5 kw
- Current: 6 A
Vessel Volume
- Extractor volume: 1.5 L
- Extraction canister volume: 1 L
- Separator 1 volume: 0.25 L
- Separators 2 & 3 volumes: 1 L
- Accumulator volume: 2.7 L
Other Specifications
- ASME certified safety valve for each vessel in isolated tubing from the vessel
- PLC microprocessor Intel i5, PLC version
- Compressed air pressure 100 Ψ
- Compressed air flow 1 CFM/h
- 1 or 2 Back pressure automatic pilot valves
- 1 or 2 Automatic back pressure valves
Configurations
This machine is composed by one or two pumps, one for CO2 and another one for co-solvent, one extraction vessels, two/three separation vessels, and a separated chiller and heater. It can reach max pressure (Pmax) of 350 bar and max temperature (Tmax) of 85 °C. There are two pressure set-points: the first one is for extraction vessels and the second one for separation vessels.
There are 5 pressure gauges and voltage can be chosen on request.
Carbon dioxide is continually recycled with a closed circuit for fluid recirculation. The system is connected to the internet by an Ethernet network connection, it’s fully automatic and is equipped with a touchscreen. PLC on board in PLC version.
The system is PED/CE certified or ASME/UL certified.
There are 5 pressure gauges and voltage can be chosen on request.
Carbon dioxide is continually recycled with a closed circuit for fluid recirculation. PLC on board in PLC version, the system is connected to the internet by an Ethernet network connection, it’s fully automatic and equipped with a touchscreen.
The system is PED/CE certified and ASME/UL certified.
Pumps
There are two pumps:
- The most important is the CO2 pump, the carbon dioxide is the principal solvent for the extraction of apolar chemical compounds.
- The secondary pump is used for the release of polar or medium apolar co-solvents during the process – useful for the extraction of chemical compounds with different degrees of polarity.
At the condition of max-pressure, this pumps can ensure a CO2 Flow of 12 kg/h, and Co-solvent Flow of 1 L/h.
The extraction process is based on small variations of solvent’s (CO2) pressure and temperature. These critical operations are ensured by several heat exchangers filled with water, which is heated or cooled by separated, external chiller and heater.
- Chiller: Included
- Heater: Included
Safety
This system can be equipped with a safety Certified PLC in addition to process PLC controlling the position of the lids, the position of the valves and other safety sensors assuring safety operations. Improving safety of the operator is our mission. CE or UL certification for the electrical panel is available as an option. Pressure vessels are ASME-certified, and control system is SIL4-rated.
System Families
Lab Series
These systems range from 1.7L up to 12L.
Production Series
These systems range from Dual 7L up to Triple 73L
Industrial Series
These systems range from Dual 120L to a 5x580L.